7 sec Steam Test Pack Kit (with moving front integrator) Photon®

Written by

in

Product type: KPCD225‑C Steam Process Challenge Device (PCD) Kit

Main application: Routine load release (especially implants), routine sterilizer monitoring, and periodic validation after installation, repair, or relocation; detects inadequate air removal and steam penetration in pre‑vacuum and gravity displacement autoclaves (132–135 °C)

Key benefit: Provides a complete challenge system combining biological and chemical indicators for robust validation of steam sterilization effectiveness

Technology / differentiator:

  • PCD225‑C porous card system containing: . Self‑contained Biological Indicator Bionova® BT225 (>10⁶ Geobacillus stearothermophilus ATCC® 7953 spores) . Type 5 chemical integrator Integron® IT26‑C (moving‑front integrator, calibrated against BIER kill time) . Self‑adhesive record card for cycle documentation
  • External Type 1 process indicator (blue → dark grey/black)
  • Instant fluorescence readout (7 seconds) for BT225, with optional visual confirmation after incubation (purple = sterilized, yellow = growth)
  • Eco‑friendly, 100% toxic heavy‑metal free formulation

Compatibility / usage environment: Designed exclusively for steam sterilization; not suitable for EO, dry heat, or radiation; ideal for healthcare, food, pharmaceutical and medical device industries