Product type: PCD224‑C Steam Test Pack Kit — Process Challenge Device (PCD) for steam sterilization
Main application: Routine release of steam‑sterilized loads, especially implants; routine sterilizer monitoring; periodic validation after installation, repair, or relocation
Key benefit: Detects inadequate air removal and steam penetration in pre‑vacuum and gravity displacement autoclaves (132–135 °C), ensuring reliable sterilization of critical loads
Technology / differentiator:
PCD224‑C porous card system containing:
Bionova® BT224 Ultra Rapid SCBI (≥10⁶ Geobacillus stearothermophilus ATCC 7953 spores per vial)
Integron® IT26‑C Type 5 moving‑front chemical integrator with clear ACCEPT/REJECT zones for easy interpretation
PCD224‑C‑RC self‑adhesive Record Card for sterilization cycle documentation
External Process Indicator (Type 1): light blue → grey
SCBI label with steam‑reactive ink: pink → brown
Compatibility / usage environment:
For steam sterilization only; suitable for healthcare, food, pharmaceutical, and medical industries